Clinical Utility of CT-based Attenuation-correction in Myocardial Perfusion SPECT Imaging
PDF
Cite
Share
Request
Original Article
P: 138-145
June 2023

Clinical Utility of CT-based Attenuation-correction in Myocardial Perfusion SPECT Imaging

Mol Imaging Radionucl Ther 2023;32(2):138-145
1. İzmir Atatürk Training and Research Hospital, Clinic of Nuclear Medicine, İzmir, Turkey
2. Ümraniye Training and Research Hospital, Clinic of Nuclear Medicine, İstanbul, Turkey
No information available.
No information available
Received Date: 16.05.2022
Accepted Date: 07.08.2022
Publish Date: 20.06.2023
PDF
Cite
Share
Request

ABSTRACT

Conclusion:

CT-based AC did not significantly contribute to diagnostic performance for increased specificity for the RCA and reduced specificity in the LAD region. Therefore, AC images should always be evaluated side by side with NAC images to benefit from the different advantages of both techniques.

Results:

Specificity, sensitivity, and accuracy were calculated as 66%, 61%, 71%, 79% and 69%, 70% for AC and NAC imaging in the whole group, respectively. There was no statistically significant difference between AC and NAC images for specificity, sensitivity, and accuracy in the male and female subgroups. In the diagnosis of right coronary artery (RCA) disease, CT AC significantly increased the specificity from 87% to 96%. However, in the left anterior descending artery (LAD) region, the specificity was significantly reduced from 95% to 77%.

Methods:

The data of 124 patients who were applied one-day stress-rest Tc-99m sestamibi SPECT/CT MPI and who had coronary angiography (CAG) results within ±3 months were retrospectively reviewed. AC and NAC images were visually evaluated by two nuclear medicine specialists in a consensus. CAG results were used as the reference standard.

Objectives:

We aimed to investigate and compare the role of computed tomography (CT)-based attenuation-corrected images (AC) with non-attenuation-corrected images (NAC) obtained by single-photon emission computed tomography/computed tomography (SPECT/CT) myocardial perfusion imaging (MPI).

Keywords:
Myocardial perfusion
scintigraphy
attenuation correction

Introduction

Myocardial perfusion imaging (MPI) is a non-invasive diagnostic test applied in the evaluation of functional severity of known or suspected coronary artery disease (CAD) (1,2). Although the diagnostic currency of MPI is accepted, attenuation developing secondary to the photon absorption caused by soft tissues such as the breast, diaphragm, thoracic wall, and prominent subdiaphragmatic gastrointestinal activity can cause the artifacts. These artifacts can ultimately affect the diagnostic performance of the test (3). Many methods such as electrocardiogram (ECG) gated imaging, prone imaging, and transmission mapping with external radioactive sources have been developed to overcome attenuation artifacts (4,5). Attenuation artifacts are partially eliminated by ECG-gated single-photon emission computed tomography (SPECT) imaging, which is frequently used in clinical practice to assess regional myocardial contractility (6). However, in inhomogeneous attenuation regions, such as thorax, attenuation maps obtained by transmission imaging are thought to be more effective than ECG-gated SPECT in attenuation correction (7). Transmission imaging conducted with external radioactive sources brings several disadvantages, such as cross-talk between emission and transmission energies and suboptimal image quality (8,9). Recently developed combined SPECT and low dose computed tomography (CT) devices can generate the transmission maps with high count and resolution in a short time, and enhance the image quality (10). Many previous studies have shown that SPECT/CT MPI has high diagnostic accuracy in the evaluation of CAD (11). Nevertheless, it increases radiation exposure, can mask some perfusion defects, and reveal false-positive defects with new unexpected artifacts (10,12). Therefore, the discussions continue on the routine usage of SPECT/CT MPI in clinical practice.

Materials and Methods

Results

Discussion

When evaluating CAD, in addition to morphological information about the location and grade of coronary artery stenosis, functional severity should be determined for the selection of the patient group that will benefit from revascularization (1,14). There are several studies showing the prognostic value of SPECT MPI used for this purpose in evaluating known or suspected CAD (15). However, the diagnostic performance of SPECT MPI is limited by low spatial resolution. The artifacts caused by soft tissue attenuation and subdiaphragmatic activity affect the diagnostic accuracy of SPECT MPI, especially by reducing specificity. Therefore,artifacts reduce the cost-effectiveness of this technique (16). Methods such as ECG-gated imaging, prone imaging, and transmission mapping have been developed for the elimination of attenuation artifacts, but all methods have some limitations. For instance, ECG-gated SPECT, which is very practical for clinical routine use, is problematic in distinguishing artifacts from ischemic tissue. While prone imaging, which is used to solve inferior wall artifacts, occasionally causes new artifacts in the anterior wall and prolongs the imaging time. Moreover, transmission maps generated by external radioactive sources, which seem to be more effective in inhomogeneous attenuation regions, also have disadvantages such as suboptimal quality transmission images and additional imaging time (8,9). In the transmission study conducted with the combined SPECT/CT system, the imaging time is not prolonged and the image quality increases due to high counting and resolution. However, new artifacts may arise and additional radiation exposure occurs (9,10).

Some studies evaluating the diagnostic efficiency of AC MPI reported that CT-AC increased specificity and accuracy significantly, while others reported that it decreased sensitivity in CAD detection with the improvement of specificity (17,18,19). In a study conducted with 120 patients, when AC images were compared with NAC imagings, the overall sensitivity decreased significantly from 87% to 70%. Overall specificity increased from 54% to 62%, although it was not statistically significant. In addition, the accuracy did not show any significant difference (12). In our study, there was no statistically significant difference between AC and NAC imaging regarding sensitivity specificity and accuracy. In the entire study group, CT-AC increased specificity from 60% to 66% and decreased sensitivity and accuracy from 79% to 71% and from 70% to 69%, respectively. However, their differences were both non-significant. In the literature, CT-AC improves the specificity in a general manner, while it results in different conclusions in terms of sensitivity and accuracy (20). This difference of outcomes may also be related to group characteristics, subjectivity affecting visual image analysis, and individual differences in interpreters, as well as technical reasons such as software used for reconstruction of the study, the use of scatter correction and resolution recovery (21,22,23). The pattern of the patients in the study group and the effect of the individual trends of the interpreters are also observed in the work of Sharma et al. (19) which used the same method and the same equipment to interpret the image.

Many factors affecting the outcome of clinical trials make it difficult to determine the diagnostic efficacy of CT-AC in CAD detection. However, as AC imaging provides high left ventricular count homogeneity, attenuation correction is thought to facilitate the separation of patients with positive and negative data according to coronary angiography. In a study conducted by Płachcińska et al. (6), specificity and accuracy in CAD detection increased statistically through AC imaging, and the increase in specificity was also observed in the male and female subgroups. According to Huang et al. (24), CT-AC reduced the defect size in the inferior wall in both male and female groups; however, AC imaging was found to be more beneficial in male than in female subgroups as diaphragmatic attenuation artifact in the inferior wall was seen more frequently in men than in women. Moreover, according to this study, attenuation correction increased the specificity for RCA disease from 77.9% to 98.7%; hence, the contribution of attenuation correction was most commonly seen in the RCA territory (24). As the current study demonstrated, the vascular area that most benefited from attenuation correction was RCA. Similar to the literature, attenuation correction yielded a decrease in the sensitivity in the whole group, while the specificity increased in the whole group and male subgroup (6).

In our study, there was no statistically significant difference in sensitivity, specificity, and accuracy between AC and NAC images in the LCx territory, as in the literature, for both whole group and subgroups. Conflicting results are reported for the LAD region (25). There are also papers that have reported that attenuation correction increases the capacitance of the detection of LAD disease (26). In contrast to Wolak et al. (27), AC imaging has also been reported to be useful in female patients (6). Additionally, some publications have argued that apical defects observed in AC images may be secondary to true apical thinning (28). It has been reported that artifacts observed in the anterior wall of NAC images were corrected in AC images (29). In the current study, however, there was no statistically significant change in sensitivity in the whole group between AC and NAC imaging in LAD region, whereas the specificity was observed to decrease in the whole group, besides the subgroups (Figure 1). In accordance with the current study, it has been reported in the literature that CT-AC reduced specificity in ​​LAD territory and it is not sufficient for the elimination of breast artifacts, as well as it has been informed that new defects occasionally arise from the apical, anterior and septal wall with AC imaging (24,30). It is known that the attenuation map (small field of view), which does not contain the entire thoracic wall, can cause these defects in oversized patients. Besides, the underestimation of the attenuation effect on the anterior thoracic wall due tomis registration of the emission and transmission images (caused by patient movement or respiratory motion) can also create new defects (31). Although re-registration seems beneficial, manual correction is based on the visual evaluation, and the correction process can vary depending on the user’s experience (32). Furthermore, regardless of misalignment, overestimation of the attenuation correction in the inferior wall, hence, a relative decrease in other segments and false perfusion defects may occur (31). All misalignments in our study were corrected manually. Nonetheless, lower specificity and accuracy values ​​were obtained in AC images in the LAD territory, and there was no statistically significant improvement in the LCx field with attenuation correction. This is thought to be caused by the effect of user-dependent realignment or overcorrection of the inferior wall, as mentioned in the literature.

Figure 1

Conclusion

In conclusion, when using CT-based attenuation correction in MPI, a visual review of raw data along with the reconstructed images is required for the correct registration of the emission and transmission data. Additionally, AC images should always be evaluated side by side with NAC images to benefit from the different advantages of both techniques.

Patients

A total of 124 patients in the age group of 35-82 (median age 61.8±11.4 years), including 71 males (57.3%) and 53 females (42.7%), who were referred to our clinic for evaluation of myocardial perfusion in 2012 and 2014 with the diagnosis of known or suspected CAD, were enrolled in our retrospective study. One-day stress-rest Tc-99m sestamibi (MIBI) ECG gated SPECT/CT imaging was applied to all subjects. All patients had available CAG data within three months before or after MPI.

Patients who were not able to perform treadmill exercise test due to their clinical condition, who had undergone pharmacological stress, and patients with left bundle branch block or arrhythmia causing insufficient counting statistics/quality in ECG-gated imaging were excluded from the study. The cases were divided into subgroups of females and males. Patients were instructed to read the detailed information form before the procedure and signed the consent form. The study protocol was approved by the Ethics Committee of Clinical Investigations of our hospital with decision number 372/2-97 dated 06.06.2016. The study was performed in accordance with the ethical standards laid down by the Declaration of Helsinki in 1964 and all its subsequent revisions. The medical history and symptoms of the patients are given in Table 1.

Table 1

Patient Preparation and SPECT/CT Myocardial Perfusion Imaging Protocol

After at least 4 h of fasting, an exercise stress test was performed in patients according to modified Bruce protocol. Calcium channel blockers or beta-blockers were stopped 48 hours before imaging and long-acting nitrates were discontinued the day before the study if there was no medical contraindication. 8-10 millicuries (mCi) [296-370 megabecquerels (MBq)] Tc-99m MIBI was injected to each patient during exercise when target heart rate was reached or an indication for the exercisetermination (physical fatigue, progressive angina, systolic blood pressure lower than 20 mmHg, systolic blood pressure higher than 250 mmHg, ST depression more than 3 mm, etc.) was observed. After the injection, patients were fed with greasy foods that accelerated the hepatobiliary clearance of the radiopharmaceutical. ECG stress images were recorded after about 45 minutes. At least 2 h after the stress test, 24-30 mCi (888-1110 MBq) of Tc-99 m MIBI injected intravenously and resting ECG-gated images were obtained after approximately 45 min following injection. Studies were conducted using a Siemens Symbia T6 (Healthcare, Erlangen, Germany) dual-head SPECT/CT hybrid camera equipped with high-resolution low energy parallel hole collimators. When the patients were in the supine position and the angle between the collimators was 90°, in a 64×64 matrix, a total of 60 projections were acquired using a step & shot technique over a 180° rotation from the 45° right anterior oblique to the 45° left posterior oblique, in 20% energy window centered on the 140 keV energy peak of Tc-99m MIBI and with a zoom factor of 1.33.

For ECG-gated SPECT imaging, all patients were monitored with a 3-lead ECG. Images were gated at 8 frames per cardiac cycle. A low-dose CT study (140 kV, 2.5 mA) of the chest was performed for attenuation mapping during both stress and rest imaging following the emission study. Obtained SPECT and AC images were automatically reconstructed using standard filtered back-projection technique and FLASH 3D iterative reconstruction algorithm, respectively. Emission and transmission data were visually inspected on axial, sagittal, and coronal fusion plans, and misalignments were manually corrected in 50 patients (40.3%).

Evaluation

Analysis of AC and NAC images was performed visually by two nuclear medicine specialists in consensus. Sections on the short, vertical, and horizontal axes obtained by the Cedars Sinai Quantitative Perfusion Gated SPECT (QPS-QGS) software packages (Cedars Sinai Medical Center, Los Angeles, CA) were evaluated.

The left ventricle was evaluated in association with the presence or absence of perfusion defects based on a 17-segment model of three main coronary arterial territories left anterior descending (LAD), left circumflex (LCx), and right coronary arteries (RCA) recommended by the American Heart Association (13).

MPI results were reported as normal in the absence of perfusion defects in both stress and rest images. If the perfusion was normal in resting images in areas where a perfusion defect was detected in stress images, ischemia (reversible defect) was reported. MPI result was reported as infarct (fixed defect) in the presence of a matched perfusion defect in both stress and rest images. However, when NAC images were reported, ECG-gated images were taken into account and areas with a fixed perfusion defect without any wall motion abnormality was interpreted as artifacts.

Coronary Angiography

CAG results of 124 patients obtained within ±3 months were analyzed. CAG findings served as the reference standard for the diagnostic performance of MPI. A reduction in the luminal diameter of ≥50% in at least one of the three coronary arteries was defined as a significant stenosis.

Statistical Analysis

All statistical analysis were performed using the IBM SPSS version 21.0 for Windows program (SPSS, Chicago, Ill.). The mean, standard deviation, minimum, and maximum values ​​of all numerical variables were calculated. NAC and AC images were compared in terms of sensitivity, specificity, and accuracy. The McNemar test was used for statistical difference analysis in assessing the diagnostic accuracy (sensitivity, specificity, accuracy) obtained by NAC and AC imaging. Results with a p value of less than 0.05 were regarded as statistically significant.

Myocardial Perfusion Imaging

In NAC images, perfusion defects were compatible with ischemia and infarction in 48 (38.7%) and 17 (%13.7) cases, respectively. Both ischemia and infarction were detected in 10 cases (8.1%). In 8 cases (6.5%) with a fixed perfusion defect in the inferior myocardial wall, MPI results were evaluated favorably for diaphragmatic attenuation because wall motion was normal in these cases. In 41 patients (33%), MPI findings were reported as normal. When the AC images of all patients were assessed, 39 cases (31.4%) were diagnosed with ischemia, 19 cases (15.3) were diagnosed with infarction, 9 cases (7.3%) were found to have both ischemia and infarction. MPI findings were reported as normal according to AC images in 57 (46%) patients.

Coronary Angiography

There was significant stenosis in 66 cases according to CAG results. Of the cases, 9 had three vessels, 17 had two vessels, 40 had single-vessel stenosis. 58 (46.8%) patients had no critical stenosis. Table 2 summarizes the MPI results of NAC and AC images according to CAG findings.

Table 2

In the entire study group, CT increased specificity (60% and 66% for NAC and AC imagings, respectively) and decreased sensitivity and accuracy (79%, 71%, and 70% and 69% for NAC and AC imagings, respectively) (Table 3). However, the results were not statistically significant. Moreover no significant effect of AC imaging was found in the male and female subgroups (Table 3).

Table 3

In the diagnosis of RCA disease, the sensitivity decreased slightly according to AC imaging. Furthermore, for the RCA region, CT-AC increased specificity and accuracy both in the entire study population and subgroups (there was statistical significance in the study group and the male subgroup regarding specificity). In the LCx territory with AC imaging, sensitivity decreased mildly, while a slight increase was observed in terms of specificity and accuracy. However, CT had no statistically significant effect on the diagnostic performance in the entire group and subgroups of patients in the LCxterritory. In the detection of LAD artery stenosis, with the effect of CT-AC, the sensitivity and accuracy did not significantly change, while specificity significantly decreased. This reduction of specificity, which was determined by attenuation correction in the LAD region, was observed similarly in subgroups as well (Table 4).

Table 4

Study Limitations

Our study was planned retrospectively with a single centered, relatively small patient sample. Another limitation of our study is that a functional diagnostic test such as SPECT MPI for the diagnosis of CAD is compared with an anatomical method such as invasive angiography as a reference method. Moreover, we did not have a prognostic value of AC MPI in our study because there were no follow-up details of the patients. According to our own clinical experience, AC imaging contributes to the detection of perfusion defects in the inferior wall; however, it reduces specificity in the evaluation of the anterior wall. Therefore, it should be kept in mind that the inferior wall defects, which are only observed in NAC images, and anterior wall defects, which are detected only in AC images, may be artefacts. It is thought that NAC images should be evaluated primarily for the findings of anterior, anteroseptal, anterolateral walls and apex.

References

1
Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, O’Gara PT, Carabello BA, Russell RO Jr, Cerqueira MD, St John Sutton MG, DeMaria AN, Udelson JE, Kennedy JW, Verani MS, Williams KA, Antman EM, Smith SC Jr, Alpert JS, Gregoratos G, Anderson JL, Hiratzka LF, Faxon DP, Hunt SA, Fuster V, Jacobs AK, Gibbons RJ, Russell RO; American College of Cardiology; American Heart Association; American Society for Nuclear Cardiology. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging--executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). J Am Coll Cardiol 2003;42:1318-1333.
2
Nakajima K, Matsuo S, Okuyama C, Hatta T, Tsukamoto K, Nishimura S, Yamashina A, Kusuoka H, Nishimura T. Cardiac event risk in Japanese subjects estimated using gated myocardial perfusion imaging, in conjunction with diabetes mellitus and chronic kidney disease. Circ J 2012;76:168-175.
3
Corbett JR, Kritzman JN, Ficaro EP. Attenuation correction for single photon emission computed tomography myocardial perfusion imaging. Curr Cardiol Rep 2004;6:32-40.
4
4 Okkalides D. Recent international regulations: low dose-low rate radiation protection and the demise of reason. Hell J Nucl Med 2008;11:86-90.
5
Kalantari F, Rajabi H, Saghar M. Quantification and reduction of attenuation related artifacts in SPET by applying attenuation model during iterative image reconstruction: a Monte Carlo study. Hell J Nucl Med 2011;14:278-283.
6
Płachcińska A, Włodarczyk M, Kovacevic-Kuśmierek K, Bieńkiewicz M, Drożdż J, Chiżyński K, Kasprzak JD, Peruga JZ, Kuśmierek J. Diagnostic performance of myocardial perfusion single-photon emission computed tomography with attenuation correction. Kardiol Pol 2016;74:32-39.
7
Zaidi H, Hasegawa B. Determination of the attenuation map in emission tomography. J Nucl Med 2003;44:291-315.
8
Garcia EV. SPECT attenuation correction: an essential tool to realize nuclear cardiology’s manifest destiny. J Nucl Cardiol 2007;14:16-24.
9
Patton JA, Townsend DW, Hutton BF. Hybrid imaging technology: from dreams and vision to clinical devices. Semin Nucl Med 2009;39:247-263.
10
Dvorak RA, Brown RK, Corbett JR. Interpretation of SPECT/CT myocardial perfusion images: common artifacts and quality control techniques. Radiographics 2011;31:2041-2057.
11
Tonge CM, Manoharan M, Lawson RS, Shields RA, Prescott MC. Attenuation correction of myocardial SPECT studies using low resolution computed tomography images. Nucl Med Commun 2005;26:231-237.
12
Apostolopoulos DJ, Savvopoulos C. What is the benefit of CT-based attenuation correction in myocardial perfusion SPET? Hell J Nucl Med 2016;19:89-92.
13
Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS; American Heart Association Writing Group on Myocardial Segmentation and Registration for Cardiac Imaging. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 2002;105:539-542.
14
Silber S, Albertsson P, Avilés FF, Camici PG, Colombo A, Hamm C, Jørgensen E, Marco J, Nordrehaug JE, Ruzyllo W, Urban P, Stone GW, Wijns W; Task Force for Percutaneous Coronary Interventions of the European Society of Cardiology. Guidelines for percutaneous coronary interventions. The Task Force for Percutaneous Coronary Interventions of the European Society of Cardiology. Eur Heart J 2005;26:804-847.
15
Pazhenkottil AP, Ghadri JR, Nkoulou RN, Wolfrum M, Buechel RR, Küest SM, Husmann L, Herzog BA, Gaemperli O, Kaufmann PA. Improved outcome prediction by SPECT myocardial perfusion imaging after CT attenuation correction. J Nucl Med 2011;52:196-200.
16
Dvorak RA, Brown RK, Corbett JR. Interpretation of SPECT/CT myocardial perfusion images: common artifacts and quality control techniques. Radiographics 2011;31:2041-2057.
17
Ficaro EP, Fessler JA, Shreve PD, Kritzman JN, Rose PA, Corbett JR. Simultaneous transmission/emission myocardial perfusion tomography. Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93:463-473.
18
Genovesi D, Giorgetti A, Gimelli A, Kusch A, D’Aragona Tagliavia I, Casagranda M, Cannizzaro G, Giubbini R, Bertagna F, Fagioli G, Rossi M, Romeo A, Bertolaccini P, Bonini R, Marzullo P. Impact of attenuation correction and gated acquisition in SPECT myocardial perfusion imaging: results of the multicentre SPAG (SPECT Attenuation Correction vs Gated) study. Eur J Nucl Med Mol Imaging 2011;38:1890-1898.
19
Sharma P, Patel CD, Karunanithi S, Maharjan S, Malhotra A. Comparative accuracy of CT attenuation-corrected and non-attenuation-corrected SPECT myocardial perfusion imaging. Clin Nucl Med 2012;37:332-338.
20
Masood Y, Liu YH, Depuey G, Taillefer R, Araujo LI, Allen S, Delbeke D, Anstett F, Peretz A, Zito MJ, Tsatkin V, Wackers FJ. Clinical validation of SPECT attenuation correction using x-ray computed tomography-derived attenuation maps: multicenter clinical trial with angiographic correlation. J Nucl Cardiol 2005;12:676-686.
21
Links JM, Becker LC, Rigo P, Taillefer R, Hanelin L, Anstett F, Burckhardt D, Mixon L. Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: a multicenter trial. J Nucl Cardiol 2000;7:414-425.
22
Hendel RC, Berman DS, Cullom SJ, Follansbee W, Heller GV, Kiat H, Groch MW, Mahmarian JJ. Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging. Circulation 1999;99:2742-2749.
23
Johansen A, Grupe P, Veje A, Braad PE, Høilund-Carlsen PF. Scatter and attenuation correction changes interpretation of gated myocardial perfusion imaging. Eur J Nucl Med Mol Imaging. Eur J Nucl Med Mol Imag 2004;31:1152-1159.
24
Huang R, Li F, Zhao Z, Liu B, Ou X, Tian R, Li L. Hybrid SPECT/CT for attenuation correction of stress myocardial perfusion imaging. Clin Nucl Med 2011;36:344-349.
25
Ficaro EP. Should SPET attenuation correction be more widely employed in routine clinical practice? For. Eur J Nucl Med Mol Imaging 2002;29:409-412.
26
Shotwell M, Singh BM, Fortman C, Bauman BD, Lukes J, Gerson MC. Improved coronary disease detection with quantitative attenuation-corrected Tl-201 images. J Nucl Cardiol 2002;9:52-62.
27
Wolak A, Slomka PJ, Fish MB, Lorenzo S, Berman DS, Germano G. Quantitative diagnostic performance of myocardial perfusion SPECT with attenuation correction in women. J Nucl Med 2008;49:915-922.
28
Okuda K, Nakajima K, Matsuo S, Wakabayashi H, Taki J, Kinuya S. Cause of apical thinning on attenuation-corrected myocardial perfusion SPECT. Nucl Med Commun 2011;32:1033-1039.
29
Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, Courter SA, Golub RJ, Case JA, Cullom SJ. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009;16:714-725.
30
Banzo I, Pena FJ, Allende RH, Quirce R, Carril JM. Prospective clinical comparison of non-corrected and attenuation- and scatter-corrected myocardial perfusion SPECT in patients with suspicion of coronary artery disease. Nucl Med Commun 2003;24:995-1002.
31
Fricke H, Fricke E, Weise R, Kammeier A, Lindner O, Burchert W. A method to remove artifacts in attenuation-corrected myocardial perfusion SPECT Introduced by misalignment between emission scan and CT-derived attenuation maps. J Nucl Med 2004;45:1619-1625.
32
Apostolopoulos DJ, Gąsowska M, Savvopoulos CA, Skouras T, Spyridonidis T, Andrejczuk A, Vassilakos PJ. The impact of transmission-emission misregistration on the interpretation of SPET/CT myocardial perfusion studies and the value of misregistration correction. Hell J Nucl Med 2015;18:114-121.